Tanabe, Masayuki and Wu, Tai Chieh and Kobayashi, Makiko and Yang, Che Hua (2019) Fabrication of Coaxial and Confocal Transducer Based on Sol-Gel Composite Material for Optical Resolution Photoacoustic Microscopy. Diagnostics, 10 (1). p. 6. ISSN 2075-4418
diagnostics-10-00006-v2.pdf - Published Version
Download (5MB)
Abstract
We have newly developed coaxial and confocal optical-resolution photoacoustic microscopy based on sol-gel composite materials. This transducer contains a concave-shaped piezoelectric layer with a focus depth of 5 mm and a hole with a diameter of 3 mm at the center to pass a laser beam into a phantom. Therefore, this system can directly detect an excited photoacoustic signal without prisms or acoustic lenses. We demonstrate the capability of the system through pulse-echo and photoacoustic imaging experiments. The center frequency of the fabricated transducer is approximately 7 MHz, and its relative bandwidth is 86%. An ex-vivo experiment is conducted, and photoacoustic signals are clearly obtained. As a result, 2- and 3-dimensional maximum amplitude projection images are reconstructed.
Item Type: | Article |
---|---|
Subjects: | STM Repository > Medical Science |
Depositing User: | Managing Editor |
Date Deposited: | 05 Jan 2023 07:16 |
Last Modified: | 08 Jun 2024 07:43 |
URI: | http://classical.goforpromo.com/id/eprint/2684 |